The Televac® MP7FR Active UHV Gauge utilizes an easily replaceable Televac® 7F double inverted magnetron cold cathode vacuum sensor. It has a variety of features including a wide range of measurement from 1*10-11 Torr to 1*10-2 Torr, a programmable set point, a factory settable analog output, and an LED display. This unit indirectly measures absolute pressure by ionizing gas in the presence of a magnetic field. Unlike hot ion sensors, this unit is resistant to inrushes of gas. The design allows for easy disassembly for replacement of the sensor.
| Operating Range | 1*10-11 Torr to 1*10-2 Torr |
| Communications | N/A |
| Analog Output | 0 V to 10 V DC |
| Programmable Set Points | 1 |
| Set Point Type | Relay |
| Supply Voltage | 24 V DC |
| Maximum Power | 6 W |
| Calibration Medium | Dry air or nitrogen |
| Overpressure | 150 psi |
| Digital Output Resolution | 2 significant digits with exponent |
| Operating Temperature | 0 °C to 50 °C |
| Storage Temperature | -20 °C to 60 °C |
| Bakeout Temperature | 150 °C (electronics removed) 400 °C (electronics and magnets removed) |
| Display Readable Distance | 3 m (10 ft) |
| Accuracy | |
| 1*10-7 Torr to 1*10-2 Torr | ±0.2 decade |
| Analog Output | ±5 mV |
| Heat treating and vacuum furnaces | |
| Cryogenics and industrial gases | |
| Thin film deposition and coating processes | |
| Semiconductor | |
| Freeze drying and food packaging | |
| Load-lock | |
| Pharmaceutical product processes |