Televac® AN 3018: Configuring the MX200 EthernetIP Address Using the MX200 Web Interface

1 Description

Purpose

This application note explains how to configure the IP address of a Televac® MX200 EthernetIP interface using the MX200 web interface. It also explains how to temporarily configure a Windows computer for direct communication with the MX200 and how to verify EthernetIP network connectivity after setup.

Network Configuration

To communicate with a Programmable Logic Controller (PLC), the MX200 EthernetIP interface must be configured with an IP address and subnet mask appropriate for the network.

For this procedure, the Windows computer is temporarily configured to communicate with the MX200 at its default network settings. After the MX200 network settings are changed, the computer is returned to its original network configuration.

Alternate Configuration Method

This application note uses the MX200 web interface to configure the EthernetIP network settings.

If internet access is available and the HMS IPConfig Utility can be used, refer to: Televac® Application Note 3017 - Configuring the MX200 EthernetIP Address Using the HMS IPConfig Utility

Additional Information

Refer to the MX200 Instruction Manual for detailed information about Class 1 and Class 3 EthernetIP communications and the complete ADI table.

After configuring the MX200 EthernetIP network settings, refer to Televac® Application Note 3016 - Loading EDS Files for Rockwell Automation Allen-Bradley CompactLogix PLCs for instructions on registering the EDS file and establishing the PLC connection.

Back to Table of Contents

2 Required Equipment, Tools, and Software

Equipment/ToolPart Number/Requirement
MX200 Vacuum Controller2-7900-37 - MX200 with the EthernetIP module installed.
Windows ComputerComputer used to access the MX200 EthernetIP web interface and configure its network settings
Ethernet CableUsed to connect the MX200 EthernetIP port to the computer
Web BrowserUsed to access the MX200 EthernetIP web interface
Network InformationIP address and subnet mask appropriate for the network where the MX200 will be installed
Current Computer Network SettingsExisting IPv4 address, subnet mask, default gateway, and other applicable settings that must be restored after configuration

IMPORTANT

Before changing the computer's network settings, record the existing IPv4 configuration.

These settings must be restored after the MX200 EthernetIP interface has been configured.

The IP address assigned to the MX200 must also be appropriate for the target network and must not conflict with another device.

Back to Table of Contents

3 Configure the Computer for Direct Communication with the MX200

3.1

Connect the MX200 EthernetIP port on the back of the unit directly to the Windows computer using an Ethernet cable.

Ethernet cable connected directly between the Windows computer and MX200 EthernetIP port
3.2

Open the Windows network settings and locate the Ethernet connection connected to the MX200.

Open the Ethernet connection status, then select: Properties

Windows Ethernet connection status window with Properties highlighted
3.3

In the Ethernet Properties window, select Internet Protocol Version 4 (TCP/IPv4).

Then click: Properties

Windows Ethernet Properties window with Internet Protocol Version 4 TCP IPv4 selected
3.4

Before changing any settings, record the computer's current IPv4 configuration.

Record the applicable values, including:

  • IP Address
  • Subnet Mask
  • Default Gateway

These settings will need to be restored after the MX200 has been configured.

Windows IPv4 Properties window showing the computer's original IP address, subnet mask, and default gateway
3.5

In the Internet Protocol Version 4 (TCP/IPv4) Properties window, select Use the following IP address, then enter the following temporary network settings. Click OK to continue.

SettingValue
IP Address192.168.0.9
Subnet Mask255.255.255.0
Default Gateway192.168.0.8
Windows IPv4 Properties window configured with temporary MX200 network settings
3.6

The Windows computer and MX200 are now on the same local network.

The MX200 EthernetIP interface uses the default IP address: 192.168.0.8

The computer is temporarily configured as: 192.168.0.9

Back to Table of Contents

4 Configure the MX200 EthernetIP Address Using the Web Interface

4.1

Open a web browser and enter the default MX200 EthernetIP address: 192.168.0.8

When prompted, enter the default username and password.

SettingValue
UsernameAdmin
Passwordadmin
Browser authentication prompt for the MX200 at its default IP address
4.2

After signing in, the MX200 EthernetIP web interface should be displayed.

Select the Configuration tab.

MX200 EthernetIP web interface with the Configuration tab selected
4.3

Enter the IP Address and Subnet Mask appropriate for the network where the MX200 will be installed.

The IP address assigned to the MX200 must not already be in use by another device on the network.

IMPORTANT

Use an IP address and subnet mask appropriate for your network. Verify that the selected IP address is not already assigned to another device.

Contact your network administrator or IT department if you are unsure which settings should be used.

4.4

After entering the desired network configuration, click Save Settings to save the changes.

MX200 EthernetIP configuration page showing the saved network settings and restart message
4.5

After saving the network settings, cycle power to the MX200 so the new EthernetIP settings take effect.

Back to Table of Contents

5 Restore the Computer Network Settings

5.1

Return to the Windows Ethernet adapter settings and open:

Internet Protocol Version 4 (TCP/IPv4) Properties

Use the same process followed in Sections 3.2 and 3.3.

5.2

Restore the computer to the original IPv4 settings recorded in Section 3.4.

This may include:

  • IP Address
  • Subnet Mask
  • Default Gateway

Enter the original values exactly as they were recorded before the temporary configuration was applied.

Windows IPv4 Properties window showing the original network settings to restore
5.3

Click OK to save the restored network configuration.

The computer should now be returned to its normal network settings and ready to communicate on the target network.

Back to Table of Contents

6 Verify Network Connectivity

6.1

Open a web browser and enter the new IP address assigned to the MX200 EthernetIP interface.

When prompted, enter:

SettingValue
UsernameAdmin
Passwordadmin
Browser authentication prompt for the MX200 at its newly assigned IP address
6.2

If the MX200 is connected to the network correctly, the EthernetIP web interface should be displayed.

From the web interface, verify that the expected EthernetIP network settings are shown.

MX200 EthernetIP web interface displayed after connecting at the new IP address
6.3

Once network connectivity has been verified, the MX200 is ready to be configured for communication with the PLC.

Refer to Televac® Application Note 3016 - Loading EDS Files for Rockwell Automation Allen-Bradley CompactLogix PLCs for instructions on registering the MX200 EDS file and creating the PLC connection.

Back to Table of Contents