Televac® AN 3011: MX200 and MM200 Analog Output Comparisons

1 Description

This application note shows the MX200 Vacuum Controller analog (0 to 10 V DC) output settings that are equivalent to an MM200 Vacuum Controller.

Back to Table of Contents

2 Comparison Chart

Green highlighted values are the default MX200 settings for that vacuum gauge type.

Red highlighted values are the default MM200 settings for that vacuum gauge type.

Vacuum Gauge MM200 Code MX200 Setting High Value Low Value
1E R1EN Linear (LIN) 1.0E+03 Torr 1.0E+00 Torr
R1ELNot supported--
1F R1EN Linear (LIN) 1.0E+04 Torr 1.0E+01 Torr
R1ELNot supported--
2A R2AN Linear (LIN) 1.0E+00 Torr 1.0E-03 Torr
R2ALLogarithmic (LOG)1.0E+00 Torr1.0E-03 Torr
R2ATNot supported--
R2ACNot supported--
R2ARNot supported--
4A R4AN Linear-by-decade (L-D) 1.0E+03 Torr 1.0E-03 Torr
R4AULinear (LIN)1.0E+00 Torr1.0E-03 Torr
R4ATNot supported--
R4AL Logarithmic (LOG) 1.0E+03 Torr 1.0E-03 Torr
5A/5B/5C/5D/5E/5F R5AN Linear (LIN) Full scale Full scale * 0.001
R5ALNot supported--
7B R7BM Linear-by-decade (L-D) 1.0E-03 1.0E-07
R7BLLogarithmic (LOG)1.0E-031.0E-07
R7BENot supported--
R7BN Linear (LIN) 1.0E-03 1.0E-06
R7BANot supported--
R7BTNot supported--
7E R7EM Linear-by-decade 7E (L-D 7E) 1.0E-02 1.0E-08
R7EL Logarithmic (LOG) 1.0E-02 1.0E-11
R7ExLinear (LIN)1.0E-0x*1.0E-(x+3)*
7F/7FC/7FCS R7FM Linear-by-decade (L-D) 1.0E-02 1.0E-11
R7FL Logarithmic (LOG) 1.0E-02 1.0E-11
R7FxLinear (LIN)1.0E-0x*1.0E-(x+3)*

*Dependent on the value of x; 7E = 2≤x≤5, 7F/7FC/7FCS = 2≤x≤8.

Back to Table of Contents