Televac® AN 3000: MX Series Replacement of MP Series Active Gauges

1 Overview

The Televac® MX Series provides a modern replacement for legacy MP Series active vacuum gauges. This application note explains the differences in analog outputs, wiring, calibration, display, and set points that should be considered when replacing an MP2AR or MP4AR with an MX2A or MX4A.

Back to Table of Contents

2 MP and MX Series Comparison

The table below summarizes the primary differences between the MX2A, MX4A, MP2AR, and MP4AR.

特征 MX2A MX4A MP2AR MP4AR
范围 1*10-4至1000托尔 1*10-4至1000托尔 1*10-3至20托尔 1*10-3至 1000 托
导向

Any
(1*10-4 to 20 Torr)

Vertical
(20 to 1000 Torr)

纵向 任何 纵向
模拟输出 7个用户可选择 7个用户可选择 2套工厂设备 3个厂区
沟通 RS-485 RS-485 无 无
响应时间 < 2 seconds < 1 second < 2 seconds < 1 second
Back to Table of Contents

3 Analog Output Compatibility

MP Series gauges provide fewer analog output options than MX Series gauges. The available output modes are summarized below.

Series Available Analog Outputs
MP Series Log, Linear, Non-linear
MX Series Log, Linear 1, Linear 2, Linear 3, Linear 4, Linear by decade, Non-linear

MP2AR and MX2A Analog Output Comparison

MP2AR输出 MX2A等效输出
直线型 线性4
非线性 No direct equivalent

MP4AR and MX4A Analog Output Comparison

MP4AR输出 MX4A等效输出
日志 日志
直线型 线性4
非线性 非线性

On MP Series gauges, analog output behavior is selected using jumpers and connector pin selection. On MX Series gauges, the analog output can be configured from the front panel or through RS-485 communications.

To configure an MX gauge to match the linear output of an MP gauge, select Linear 4 using the following steps:

  1. Navigate to the Measurement Screen.
  2. Press SEL four times. The screen will read Analog Out.
  3. Press ENT to unlock the screen.
  4. Press DOWN three times. The screen will read Linear 4.
  5. Press ENT to save the change.
Back to Table of Contents

4 Wiring and Pin Connections

The MP and MX Series use different connectors and pin assignments. Existing MP wiring must therefore be updated when installing an MX replacement.

Connector Comparison

MP连接器 MX连接器
9-pin D-sub 15-pin D-sub

Pin Mapping

连接 MP引脚 MX Pin
功率(+) 引脚3 引脚8
功率(-) 引脚4 引脚15
对数/线性模拟输出(+) 引脚5 引脚9
非线性模拟输出(+) 引脚7 引脚9
模拟量输出(-) 引脚8 引脚1
设定点继电器NO 引脚1 引脚4
设定点继电器NC 引脚2 引脚3
设定点继电器C 引脚6 引脚11

Overall Pin Diagram Comparison

MP and MX active gauge pin diagram comparison
Back to Table of Contents

5 Calibration

Calibration points on MP Series gauges are adjusted using potentiometers (pots). MX Series calibration is performed through the front panel or RS-485 communications.

MP2AR和MX2A校准点

校准点 MP2AR MX2A
零 是的,是的 是的,是的
10托尔 没有 是的,是的
760托尔(大气) 没有 是的,是的

MP4AR和MX4A校准点

校准点 MP4AR MX4A
零 是的,是的 是的,是的
1托尔 没有 是的,是的
760托尔(大气) 是的,是的 是的,是的

The additional calibration points available on MX Series gauges allow calibration to be performed at more points across the measurement range.

Back to Table of Contents

6 Display

MX Series gauges use a higher-resolution front-facing OLED display, while MP Series gauges use a top-mounted 3-character LED display.

说明 MP显示 MX显示器
种类 3-character 7-segment LED 256 × 64 color OLED
场地 顶端 前面
可读距离 3米(10英尺) 3米(10英尺)

The MX display also provides access to configuration settings directly from the front panel.

Back to Table of Contents

7 Set Points

MX Series gauges provide two set points and support higher switching loads than MP Series gauges.

说明 MP MX
设置点 1 2
种类 中继 继电器,开路集电极
继电器触点等级 0.5 A @ 24 V DC 2 A @ 220 V DC (250 V AC)
开放式收藏家评级 无 2 A @ 60 V DC
Back to Table of Contents

8 RS-485 Communications

MX Series gauges include RS-485 communications, while MP Series gauges do not. RS-485 allows multiple addressable gauges to share the same communications bus and enables remote configuration without accessing the gauge's front panel.

Back to Table of Contents