1 Introduction
1.1 Disclaimer
All information in this manual is subject to change without notice. The Fredericks Company assumes no responsibility for inaccuracies in product specifications or any liability arising from product use. Please contact us online or call +1 215 947 2500 with comments or questions.
1.2 Description
The Compact 2A is a thermocouple vacuum sensor controller with a range of 1*10-3 Torr to 20 Torr that utilizes several Televac® thermocouple vacuum sensors including the 2A Thermocouple, 2A VacuMini, and 2A NASA Mini vacuum gauges. The Compact 2A has a 0-10 Volt analog output.
1.3 Operating Specifications
| 工作范围 | 1*10-3托尔至20托尔。 |
| 沟通 | NA |
| 可编程设置点 | NA |
| 模拟输出 | 0 to 10 V Non-linear |
| 电源电压 | 15 V AC, 50-60 Hz (230 V Optional) |
| 校准介质 | 干燥空气或氮气 |
| 过压 | 依赖传感器 |
| 模拟输出分辨率 | 12 位 |
| 操作温度 | 0℃至50℃ |
| 储存温度 | -20℃至60℃ |
| 烘烤温度 | 依赖传感器 |
| 响应时间 | ≤ 1 second |
| 准确度 | |
| 1 mTorr至10 mTorr | ±1 mTorr |
| 10 mTorr to 500 mTorr | ±10% |
| 500 mTorr to 1 Torr | ±20% |
| 模拟输出 | ±10 mV |
| 显示可读距离 | 1 米(3 英尺) |
1.4 Safety Information
START BY READING THESE IMPORTANT SAFETY INSTRUCTIONS AND NOTES
In these instructions the word “product” refers to the Compact 2A and all of its approved parts and accessories.
NOTE: These instructions do not and cannot provide for every contingency that may arise in connection with the installation, operation, or maintenance of this product. Should you require further assistance, please contact us.
This product has been designed and tested to offer reasonably safe service provided it is installed, operated and serviced in strict accordance with these safety instructions.
These safety precautions must be observed during all phases of operation, installation, and service of this product. Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards of design, manufacture, and intended use of the instrument. Televac® disclaims all liability for the customer’s failure to comply with these requirements.
- Read all safety and operating instructions before use.
- Retain instructions for future reference.
- Heed all warnings on the product and in the operating instructions.
- Follow all operating and maintenance instructions.
- Accessories: Do not use accessories not recommended in this manual.
The Compact 2A qualifies as a Safety Extra-Low Voltage (SELV) device. As such, it represents little to no hazard concerning electrical shock or burns.
Do not substitute parts or modify instrument. Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modifications to the product. Return the product to Televac for service and repair to ensure that safety features are maintained. Do not use this product if it has unauthorized modifications.
1.5 Dimensions

2. Setup
2.1 Installation
Each Compact 2A is designed to be used in conjunction with a 2A Thermocouple sensor. The 2A sensor can be mounted in any orientation. The sensor should be mounted close to the area where vacuum measurement is desired. Each 2A sensor has a key that only allows the sensor end of the Compact 2A cable to mate with the sensor in the proper orientation. Rotate the sensor end of the Compact 2A cable until the correct alignment is obtained and the Compact 2A cable is able to slide onto the 2A sensor. Connect the RJ14 end of the Compact 2A cable into the port on the back of the Compact 2A. It is recommended that the ground connector on the cable is also connected to the port on the back of the Compact 2A labeled with the universal earth ground symbol.
Try to avoid connecting the 2A sensor to the vacuum chamber by long or narrow piping as this may affect the accuracy and response time of the unit. Avoid mounting the unit near a heater within the chamber as this may affect the measurement accuracy. Excessive vibration of the sensor may affect accuracy and decrease the life of the sensor. Exposure to oils and other contaminants will decrease the accuracy and decrease the life of the sensor. Do not expose the sensor to corrosive gases.
2.2 Electrical Information
The Compact 2A has several connectors located on the rear of the unit. A connection diagram is provided for convenience.
| 标签 | 说明 |
| 1 | Thermocouple Sensor Input |
| 2 | AC Input |
| 3 | 模拟输出 |
| 4 | Thermocouple Sensor Ground |

2.3 Calibration
The Compact 2A controller provides a calibration adjustment to maximize the accuracy of each sensor. The adjustment should be set at 1 mTorr. Maximum accuracy will utilize a 2A sensor and a NIST traceable reference gauge co-located on a vacuum chamber. A controlled leak of dry nitrogen is necessary for calibration. All Compact 2A controllers purchased with a 2A sensor will be factory calibrated and there is no need for adjustment.
For calibration, pump the vacuum chamber down to below 1*10-4 Torr. Using the leak control on the vacuum system, adjust the pressure in the chamber until the reference gauge reads 1 mTorr. Then adjust the potentiometer located in the rear of the unit until the display shows 1 mTorr.
2.4 Analog Output
The Compact 2A contains a 0-10 V Non-Linear analog output for use with voltmeters, programmable logic controllers or other data acquisition devices. To convert the analog output signal to a pressure, refer to Figure 3.

3 Part Numbers
The following table lists available part numbers for the Televac® Compact 2A Vacuum Controller and accessories.
小型2A真空控制器
| 说明 | 部件号 | |
| 小型2A真空控制器 | 2-3014-042 | Buy Now |
2A模块化连接器电缆
2A型迷你模块连接器电缆
| 说明 | 部件号 | |
| 10′ 2A Mini Modular Connector Cable | 2-9800-096 | Buy Now |
| 10′ Coiled 2A Mini Modular Connector Cable | 2-9800-082 | Buy Now |
备件
| 说明 | 部件号 | |
| Power Cord | 1-9800-49 | Buy Now |