Part Numbers: 2-7980-031, 2-7980-032, 2-7980-052
Caratteristiche e vantaggi:
- Manometro per vuoto ad ampio raggio molto compatto
- Display digitale a LED di facile lettura del vuotometro
- Eccellente risoluzione e ripetibilità del vuotometro
- Prestazioni superiori da unità a unità tra i vuotometri
- ISO 17025 accredited and NIST traceable calibrations optional
- Eccellente assistenza clienti per i vuotometri
The Televac® CC-10 Wide Range Active Vacuum Gauge (also called an analog vacuum gauge, digital vacuum gauge, or vacuum pressure gauge) utilizes two internal sensors to measure full range vacuum from UHV (10-9 Torr) to above atmosphere (103 or 1000 Torr). This vacuum gauge has a variety of features including RS-485 communications, three programmable set points, and a selectable analog vacuum pressure gauge output.
Il luminoso display digitale a LED del vuotometro rende facile la lettura a distanza. Le opzioni di questo manometro per vuoto ad ampia portata possono essere impostate da quattro comandi a sfioramento sulla parte anteriore dell'unità o attraverso la comunicazione RS-485. Questo manometro per vuoto offre un'ampia selezione di raccordi, un uso semplice e una facile sostituzione del sensore che lo rendono una scelta eccellente per una varietà di applicazioni per il vuoto.
CC-10 Misuratore di vuoto attivo ad ampio raggio
D-Sub 15, 15 Conductor Cable (Flying Leads)
Alimentazione a parete
Pezzi di ricambio
Campo di funzionamento | 1*10-9 Torr a 1000 Torr |
Comunicazioni | RS-485 |
Uscita analogica | 0 V a 10 V lineare/logaritmico |
Set point programmabili | 3 relè |
Tensione di alimentazione | 22 V DC to 26 V DC |
Potenza massima | 8 W |
Mezzo di calibrazione | Aria secca o azoto |
Sovrappressione | 150 PSI |
Risoluzione di uscita digitale | 2 cifre significative con esponente |
Risoluzione di uscita analogica | 16 bit |
Temperatura di esercizio | 0 °C a 50 °C |
Temperatura di stoccaggio | da -40 °C a 85 °C |
Temperatura di Bakeout | 110 °C (electronics removed) |
Tempo di risposta | ≤500 ms |
Display Readable Display | 4 m (13 ft) |
Precisione | |
10-9 Torr to 10-1 Torr | ±0,2 di decade |
10-1 Torr to 200 Torr | ±15% of reading |
200 Torr to 1000 Torr | ±5% della lettura |
Uscita analogica | ±10 mV |
Pin | Descrizione |
1 | Analog Out (Ground) |
2 | RS-485 A (-) |
3 | SP3 Out |
4 | SP2 Out |
5 | SP1 Out |
6 | Zero adjust input (signal in) |
7 | EXT Vcc (signal in) |
8 | Supply Voltage (+) |
9 | Analog Out (0 to 10) V |
10 | RS-485 B (+) |
11 | SP2, SP3 COM |
12 | SP1 COM |
13 | HV inhibit signal (signal in) |
14 | Atm adjust input (signal in) |
15 | Supply Voltage (-) |
<STX>0R1<CR> | Read units |
<STX>0R2<CR> | Read SP1L and SP1H |
<STX>0R5<CR> | Read analog output set parameter |
<STX>0W1XXXX<CR> | Set pressure unit |
<STX>0W2XXXXXXXX<CR> | Set SP1L and SP1H |
<STX>0C1<CR> | Crystal gauge atm adjustment |
<STX>0S1<CR> | Read pressure data |
<STX>0S2<CR> | Read error status |
Assume address 0
Full list of commands is available in the instruction manual.
The CC-10 has three set points. All three are optically coupled solid state relays that are rated for 0.12 A at 50 V DC. Each set point has hysteresis characteristics and consists of two set values: upper threshold and lower threshold. LEDs on the front panel display the status of each set point. The high and low threshold values of each of the three set points can be set independently.
Molibdeno |
Glass |
Copper |
Stainless Steel |
Chromium |
Solder |
Fluorocarbon Elastomer |
Quarzo |
Enclosure Material | Metallized plastic |
Collegamenti elettrici | D-Sub 15-pin male |
Peso con sensore | 698 g (1.54 lbs) (NW25 Flange) |
Peso senza sensore | 355 g (0.78 lbs) |
CE certified to EN 61010-1, EN 61236-1 |
RoHS compliant |
IP40 |
ISO 17025 accredited calibration optional |
Calibrazione tracciabile NIST opzionale |
Documentazione del prodotto
CC-10 Wide Range Datasheet |
CC-10 Wide Range Instruction Manual |
Televac®️ Vacuum Calibration Tolerances |
3D Models (STEP Files)
2-7980-031 – CC-10 NW25/KF25 Stainless Steel |
2-7980-032 – CC-10 NW40/KF40 Stainless Steel |
2-7980-052 – CC-10 CF40/2.75″ CF-F Stainless Steel |
Note applicative
Televac® Application Note 3015: Recommended Practices for Vacuum Calibration |
Strumenti
Convertitore di misura del vuoto Televac |
Web Serial Terminal |
Articoli
Vacuum Measurement Using Modern Cold Cathode Technology |
Brochures
Televac®️ Vacuum Measurement Product Catalog |
CC-10 Wide Range Active Vacuum
Gauge Quantity Discounts
Quantità | Discount |
1 | List |
5 | 2% |
10 | 5% |
15 | 7% |
20 | 10% |
25 | 13% |
30+ | 15% |
D-Sub 15 Cable Quantity Discounts
Quantità | Discount |
1 | List |
50 | 5% |
75 | 10% |
100+ | 15% |