Televac® AN 3000: MX Series Replacement of MP Series Active Gauges

1 Overview

The Televac® MX Series provides a modern replacement for legacy MP Series active vacuum gauges. This application note explains the differences in analog outputs, wiring, calibration, display, and set points that should be considered when replacing an MP2AR or MP4AR with an MX2A or MX4A.

Back to Table of Contents

2 MP and MX Series Comparison

The table below summarizes the primary differences between the MX2A, MX4A, MP2AR, and MP4AR.

Caratteristica MX2A MX4A MP2AR MP4AR
Gamma 1*10-4 a 1000 Torr 1*10-4 a 1000 Torr 1*10-3 a 20 Torr 1*10-3 to 1000 Torr
Orientamento

Any
(1*10-4 to 20 Torr)

Vertical
(20 to 1000 Torr)

Verticale Qualsiasi Verticale
Uscite analogiche 7 selezionabili dall'utente 7 selezionabili dall'utente 2 set di fabbrica 3 set di fabbrica
Comunicazioni RS-485 RS-485 Nessuno Nessuno
Tempo di risposta < 2 seconds < 1 second < 2 seconds < 1 second
Back to Table of Contents

3 Analog Output Compatibility

MP Series gauges provide fewer analog output options than MX Series gauges. The available output modes are summarized below.

Series Available Analog Outputs
MP Series Log, Linear, Non-linear
MX Series Log, Linear 1, Linear 2, Linear 3, Linear 4, Linear by decade, Non-linear

MP2AR and MX2A Analog Output Comparison

Uscita MP2AR Uscita equivalente MX2A
Lineare Lineare 4
Non lineare No direct equivalent

MP4AR and MX4A Analog Output Comparison

Uscita MP4AR Uscita equivalente MX4A
Log Log
Lineare Lineare 4
Non lineare Non lineare

On MP Series gauges, analog output behavior is selected using jumpers and connector pin selection. On MX Series gauges, the analog output can be configured from the front panel or through RS-485 communications.

To configure an MX gauge to match the linear output of an MP gauge, select Linear 4 using the following steps:

  1. Navigate to the Measurement Screen.
  2. Press SEL four times. The screen will read Analog Out.
  3. Press ENT to unlock the screen.
  4. Press DOWN three times. The screen will read Linear 4.
  5. Press ENT to save the change.
Back to Table of Contents

4 Wiring and Pin Connections

The MP and MX Series use different connectors and pin assignments. Existing MP wiring must therefore be updated when installing an MX replacement.

Connector Comparison

Connettore MP Connettore MX
9-pin D-sub 15-pin D-sub

Pin Mapping

Collegamento MP Pin MX Pin
Potenza (+) Pin 3 Pin 8
Potenza (-) Pin 4 Pin 15
Log/uscita analogica lineare (+) Pin 5 Pin 9
Uscita analogica non lineare (+) Pin 7 Pin 9
Uscita analogica (-) Pin 8 Pin 1
Set Point Relè NO Pin 1 Pin 4
Set Point Relè NC Pin 2 Pin 3
Set Point Relè C Pin 6 Pin 11

Overall Pin Diagram Comparison

MP and MX active gauge pin diagram comparison
Back to Table of Contents

5 Calibration

Calibration points on MP Series gauges are adjusted using potentiometers (pots). MX Series calibration is performed through the front panel or RS-485 communications.

Punti di taratura MP2AR e MX2A

Punto di calibrazione MP2AR MX2A
Zero Sì Sì
10 Torr No Sì
760 Torr (Atmosfera) No Sì

Punti di taratura MP4AR e MX4A

Punto di calibrazione MP4AR MX4A
Zero Sì Sì
1 Torr No Sì
760 Torr (Atmosfera) Sì Sì

The additional calibration points available on MX Series gauges allow calibration to be performed at more points across the measurement range.

Back to Table of Contents

6 Display

MX Series gauges use a higher-resolution front-facing OLED display, while MP Series gauges use a top-mounted 3-character LED display.

Descrizione Display MP Display MX
Tipo 3-character 7-segment LED 256 × 64 color OLED
Posizione Top Parte anteriore
Distanza leggibile 3 m (10 piedi) 3 m (10 piedi)

The MX display also provides access to configuration settings directly from the front panel.

Back to Table of Contents

7 Set Points

MX Series gauges provide two set points and support higher switching loads than MP Series gauges.

Descrizione MP MX
Punti fissi 1 2
Tipo Relè Relè, collettore aperto
Valutazione del contatto del relè 0.5 A @ 24 V DC 2 A @ 220 V DC (250 V AC)
Valutazione del collettore aperto Nessuno 2 A @ 60 V DC
Back to Table of Contents

8 RS-485 Communications

MX Series gauges include RS-485 communications, while MP Series gauges do not. RS-485 allows multiple addressable gauges to share the same communications bus and enables remote configuration without accessing the gauge's front panel.