Televac® AN 3004: Cleaning the 7FC and 7FCS Cold Cathode Passive Gauges

1 Description

This application note explains the process for cleaning the Televac® 7FC Cold Cathode Vacuum Gauge and the 7FCS Cold Cathode Vacuum Gauge that have been used in harsh environments or operating for extended periods in the 10-4 Torr or higher range.

Back to Table of Contents

2 Materials Needed

The following materials are needed to properly clean the 7FC and 7FCS Cold Cathode Vacuum Gauges.

  • 3/16" Allen Wrench
  • Scouring pad
Back to Table of Contents

3 Instructions

Watch the demonstration below for a step-by-step example of cleaning the 7FC and 7FCS Cold Cathode Vacuum Gauges. The complete text instructions are available below the video.

View Text Instructions
3.1

Remove the vacuum gauge from the vacuum chamber.

3.2

Unscrew the bolts to expose the anode assembly.

3.3

Gently lift the anode assembly upwards to separate it from the body.

3.4

Remove the copper gasket from the vacuum gauge.

3.5

Check the knife-edge on both the anode and the body to make sure they are in good condition.

3.6

Gently remove any contaminants from the anode using a scouring pad.

For the 7FCS, be careful not to damage the starting aid.

3.7

Blast the body with glass beads.

3.8

After cleaning, blow out any residual glass beads or dust with dry nitrogen.

3.9

Install a new copper gasket in the body.

3.10

Reinsert the anode assembly into the body.

3.11

Tighten the bolts gradually, moving from bolt to bolt to ensure uniform compression of the gasket.

3.12

Reinstall on the vacuum chamber.

3.13

Allow several hours for the tube to degas when pumped down to high vacuum.

3.14

If properly cleaned and assembled, the gauge tube is ready for use without recalibration unless NIST traceable or ISO 17025 accredited calibration is required.