1. Description
General
This document gives a description of the calibration options available from Televac® as well as the vacuum calibration test points and tolerances for various vacuum sensors, vacuum controllers, and active vacuum gauges offered by Televac®. All values are in Torr and calibrations are done in nitrogen gas unless otherwise stated. Televac® reserves the right to change vacuum test points and tolerances for any product. Any modification to any unit without approval from Televac® voids the unit’s calibration certificate or certificate of compliance. Additional custom test points are available upon request.
ISO/IEC 17025 Accredited Calibration
The Fredericks Company is ISO/IEC 17025 accredited for vacuum and pressure calibration in the range of 1E-6 Torr to 1000 Torr. Units are vacuum calibrated (on a vacuum stand) in nitrogen gas at various points (gauge dependent) against reference standards traceable to NIST, using ISO/IEC 17025 accredited systems. Units of measurement of Torr are used for all calibrations. This includes incoming (as found) data, outgoing (as left) data after calibration adjustments, the reference standard uncertainty at each point, and a statement of conformity.
Additional Notes on ISO/IEC 17025 Accredited Calibration
The Fredericks Company’s ISO/IEC 17025 accredited vacuum calibration certificates, unless explicitly requested otherwise, will by default include a statement of conformity using a simple decision rule, will by default use the order acknowledgement “ship to” address as the customer name and address, and will by default specify a 1 year recommended calibration interval.
Calibrazione con tracciabilità NIST
The Fredericks Company offers NIST traceable calibration, where units are vacuum calibrated (on a vacuum stand) in nitrogen gas at various points (gauge dependent) against reference standards traceable to NIST. This includes incoming (as found) data, outgoing (as left) data after calibration adjustments, but does not include the reference standard uncertainty at each point (ISO 17025 accredited calibration does include uncertainties).
Calibrazione di fabbrica
The Fredericks Company offers factory calibration, where units are electrically verified using electrical equipment only. The verification includes checking that the units provide vacuum readings within a specified tolerance when a specific electrical signal is applied. This is not a vacuum calibration, and does not guarantee that a sensor will read accurately when connected to a controller or active gauge. This is because sensor contamination and sensor to sensor mechanical variability cannot be accounted for with an electrical verification.
Back to Table of Contents2. Comparison Table of Calibration Options
| Descrizione | Fabbrica | Tracciabile NIST | Accreditato ISO 17025 |
|---|---|---|---|
| Verifica elettrica | Sì | No | No |
| Calibrazione di confronto a vuoto con riferimenti tracciabili NIST | No | Sì | Sì |
| Incertezze dello standard di riferimento incluse | No | No | Sì |
| Taratura accreditata ISO 17025 | No | No | Sì |
3. 2A, 2A VacuMini, 2A NASA Mini Part 1
Applicable Controllers and Active Gauges
MX200, MM200, MC300, MV2A, Compact 2A, Vacuguard, B2A Portable, MP2AR
Calibrazione di fabbrica
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 0.0E+0 | -1.00E-3 | 1.00E-3 | ±1.00E-3 |
| 1.0E-1 | 9.00E-2 | 1.10E-1 | ±10% |
| 1.0E+0 | 8.00E-1 | 1.20E+0 | ±20% |
Calibrazione con tracciabilità NIST
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 0.0E+0 | -1.00E-3 | 1.00E-3 | ±1.00E-3 |
| 1.0E-2 | 9.00E-3 | 1.10E-2 | ±10% |
| 2.0E-2 | 1.80E-2 | 2.20E-2 | ±10% |
| 5.0E-2 | 4.50E-2 | 5.50E-2 | ±10% |
| 1.0E-1 | 9.00E-2 | 1.10E-1 | ±10% |
| 2.0E-1 | 1.60E-1 | 2.40E-1 | ±20% |
| 5.0E-1 | 4.00E-1 | 6.00E-1 | ±20% |
| 1.0E+0 | 8.00E-1 | 1.20E+0 | ±20% |
Calibrazione accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0 | 2.00E-3 | ±1.00E-3 or ±100% |
| 1.0E-2 | 9.00E-3 | 1.10E-2 | ±10% |
| 2.0E-2 | 1.80E-2 | 2.20E-2 | ±10% |
| 5.0E-2 | 4.50E-2 | 5.50E-2 | ±10% |
| 1.0E-1 | 9.00E-2 | 1.10E-1 | ±10% |
| 2.0E-1 | 1.60E-1 | 2.40E-1 | ±20% |
| 5.0E-1 | 4.00E-1 | 6.00E-1 | ±20% |
| 1.0E+0 | 8.00E-1 | 1.20E+0 | ±20% |
Calibrazione compatibile con la specifica GE Aviation P10TF3, accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0* | 2.81E-3 | ±0.45 decade or +182%, -65% |
| 1.0E-2 | 5.00E-3 | 1.50E-2 | ±5.00E-3 or ±50% |
| 2.0E-2 | 1.50E-2 | 2.50E-2 | ±5.00E-3 or ±25% |
| 5.0E-2 | 4.25E-2 | 5.75E-2 | ±15% |
| 1.0E-1 | 8.50E-2 | 1.15E-1 | ±15% |
| 2.0E-1 | 1.70E-1 | 2.30E-1 | ±15% |
| 5.0E-1 | 4.25E-1 | 5.75E-1 | ±15% |
| 1.0E+0 | 8.50E-1 | 1.15E+0 | ±15% |
ISO 17025 Accredited AMS 2769 Compatible Calibration
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0* | 5.00E-3 | ±4.00E-3 or ±400% |
| 1.0E-2 | 6.00E-3 | 1.40E-2 | ±40% |
| 2.0E-2 | 1.20E-2 | 2.80E-2 | ±40% |
| 5.0E-2 | 3.00E-2 | 7.00E-2 | ±40% |
| 1.0E-1 | 6.00E-2 | 1.40E-1 | ±40% |
| 2.0E-1 | 1.20E-1 | 2.80E-1 | ±40% |
| 5.0E-1 | 3.00E-1 | 7.00E-1 | ±40% |
| 1.0E+0 | 6.00E-1 | 1.40E+0 | ±40% |
Note: The calculated limit is negative, which isn't possible for absolute pressure, or is outside the gauge's measurement or display range, so the limit shown is the gauge's minimum or maximum reading.
4. 2A, 2A VacuMini, 2A NASA Mini Part 2
Applicable Controllers and Active Gauges
Calibrazione di fabbrica
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 0.0E+0 | 0.00E+0* | 1.00E-3 | ±1.00E-3 |
| 1.0E+0 | 7.00E-1 | 1.30E+0 | ±30% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
Calibrazione con tracciabilità NIST
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 0.0E+0 | 0.00E+0* | 1.00E-3 | ±1.00E-3 |
| 1.0E-2 | 9.00E-3 | 1.10E-2 | ±10% |
| 2.0E-2 | 1.80E-2 | 2.20E-2 | ±10% |
| 5.0E-2 | 4.50E-2 | 5.50E-2 | ±10% |
| 1.0E-1 | 9.00E-2 | 1.10E-1 | ±10% |
| 2.0E-1 | 1.80E-1 | 2.20E-1 | ±10% |
| 5.0E-1 | 4.50E-1 | 5.50E-1 | ±10% |
| 1.0E+0 | 7.00E-1 | 1.30E+0 | ±30% |
Calibrazione accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0 | 2.00E-3 | ±1.00E-3 or ±100% |
| 1.0E-2 | 9.00E-3 | 1.10E-2 | ±10% |
| 5.0E-2 | 4.50E-2 | 5.50E-2 | ±10% |
| 1.0E-1 | 9.00E-2 | 1.10E-1 | ±10% |
| 5.0E-1 | 4.50E-1 | 5.50E-1 | ±10% |
| 1.0E+0 | 7.00E-1 | 1.30E+0 | ±30% |
| 1.0E+1 | 7.00E+0 | 1.30E+1 | ±30% |
| 1.0E+2 | 7.00E+1 | 1.30E+2 | ±30% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
Calibrazione compatibile con la specifica GE Aviation P10TF3, accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0* | 2.81E-3 | ±0.45 decade or +182%, -65% |
| 1.0E-2 | 5.00E-3 | 1.50E-2 | ±5.00E-3 or ±50% |
| 5.0E-2 | 4.25E-2 | 5.75E-2 | ±15% |
| 1.0E-1 | 8.50E-2 | 1.15E-1 | ±15% |
| 5.0E-1 | 4.25E-1 | 5.75E-1 | ±15% |
| 1.0E+0 | 8.50E-1 | 1.15E+0 | ±15% |
| 1.0E+1 | 9.00E+0 | 1.10E+1 | ±10% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
ISO 17025 Accredited AMS 2769 Compatible Calibration
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0* | 5.00E-3 | ±4.00E-3 or ±400% |
| 1.0E-2 | 6.00E-3 | 1.40E-2 | ±40% |
| 5.0E-2 | 3.00E-2 | 7.00E-2 | ±40% |
| 1.0E-1 | 6.00E-2 | 1.40E-1 | ±40% |
| 5.0E-1 | 3.00E-1 | 7.00E-1 | ±40% |
| 1.0E+0 | 6.00E-1 | 1.40E+0 | ±40% |
| 1.0E+1 | 6.00E+0 | 1.40E+1 | ±40% |
| 1.0E+2 | 6.00E+1 | 1.40E+2 | ±40% |
| 7.6E+2 | 4.56E+2 | 1.00E+3* | ±40% |
Note: The calculated limit is negative, which isn't possible for absolute pressure, or is outside the gauge's measurement or display range, so the limit shown is the gauge's minimum or maximum reading.
5. 4A Part 1
Applicable Controllers and Active Gauges
Calibrazione di fabbrica
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 0.0E+0 | 0.00E+0* | 1.00E-3 | ±1.00E-3 |
| 1.0E+0 | 9.00E-1 | 1.10E+0 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
Calibrazione con tracciabilità NIST
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 0.0E+0 | 0.00E+0* | 1.00E-3 | ±1.00E-3 |
| 1.0E-2 | 9.00E-3 | 1.10E-2 | ±10% |
| 5.0E-2 | 4.50E-2 | 5.50E-2 | ±10% |
| 1.0E-1 | 9.00E-2 | 1.10E-1 | ±10% |
| 5.0E-1 | 4.50E-1 | 5.50E-1 | ±10% |
| 1.0E+0 | 9.00E-1 | 1.10E+0 | ±10% |
| 1.0E+1 | 9.00E+0 | 1.10E+1 | ±10% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
Calibrazione accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0 | 2.00E-3 | ±1.00E-3 or ±100% |
| 1.0E-2 | 9.00E-3 | 1.10E-2 | ±10% |
| 5.0E-2 | 4.50E-2 | 5.50E-2 | ±10% |
| 1.0E-1 | 9.00E-2 | 1.10E-1 | ±10% |
| 5.0E-1 | 4.50E-1 | 5.50E-1 | ±10% |
| 1.0E+0 | 9.00E-1 | 1.10E+0 | ±10% |
| 1.0E+1 | 9.00E+0 | 1.10E+1 | ±10% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
Calibrazione compatibile con la specifica GE Aviation P10TF3, accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0* | 2.81E-3 | ±0.45 decade or +182%, -65% |
| 1.0E-2 | 5.00E-3 | 1.50E-2 | ±5.00E-3 or ±50% |
| 5.0E-2 | 4.25E-2 | 5.75E-2 | ±15% |
| 1.0E-1 | 8.50E-2 | 1.15E-1 | ±15% |
| 5.0E-1 | 4.25E-1 | 5.75E-1 | ±15% |
| 1.0E+0 | 8.50E-1 | 1.15E+0 | ±15% |
| 1.0E+1 | 9.00E+0 | 1.10E+1 | ±10% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
ISO 17025 Accredited AMS 2769 Compatible Calibration
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0* | 5.00E-3 | ±4.00E-3 or ±400% |
| 1.0E-2 | 6.00E-3 | 1.40E-2 | ±40% |
| 5.0E-2 | 3.00E-2 | 7.00E-2 | ±40% |
| 1.0E-1 | 6.00E-2 | 1.40E-1 | ±40% |
| 5.0E-1 | 3.00E-1 | 7.00E-1 | ±40% |
| 1.0E+0 | 6.00E-1 | 1.40E+0 | ±40% |
| 1.0E+1 | 6.00E+0 | 1.40E+1 | ±40% |
| 1.0E+2 | 6.00E+1 | 1.40E+2 | ±40% |
| 7.6E+2 | 4.56E+2 | 1.00E+3* | ±40% |
Note: The calculated limit is negative, which isn't possible for absolute pressure, or is outside the gauge's measurement or display range, so the limit shown is the gauge's minimum or maximum reading.
6. 4A Part 2
Applicable Controllers and Active Gauges
Calibrazione di fabbrica
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 0.0E+0 | -1.00E-3 | 1.00E-3 | ±1.00E-3 |
| 1.0E+0 | 9.00E-1 | 1.10E+0 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
Calibrazione con tracciabilità NIST
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 0.0E+0 | -1.00E-3 | 1.00E-3 | ±1.00E-3 |
| 1.0E-2 | 9.00E-3 | 1.10E-2 | ±10% |
| 5.0E-2 | 4.50E-2 | 5.50E-2 | ±10% |
| 1.0E-1 | 9.00E-2 | 1.10E-1 | ±10% |
| 5.0E-1 | 4.50E-1 | 5.50E-1 | ±10% |
| 1.0E+0 | 9.00E-1 | 1.10E+0 | ±10% |
| 1.0E+1 | 8.00E+0 | 1.20E+1 | ±20% |
| 1.0E+2 | 8.00E+1 | 1.20E+2 | ±20% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
Calibrazione accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0 | 2.00E-3 | ±1.00E-3 or ±100% |
| 1.0E-2 | 9.00E-3 | 1.10E-2 | ±10% |
| 5.0E-2 | 4.50E-2 | 5.50E-2 | ±10% |
| 1.0E-1 | 9.00E-2 | 1.10E-1 | ±10% |
| 5.0E-1 | 4.50E-1 | 5.50E-1 | ±10% |
| 1.0E+0 | 9.00E-1 | 1.10E+0 | ±10% |
| 1.0E+1 | 8.00E+0 | 1.20E+1 | ±20% |
| 1.0E+2 | 8.00E+1 | 1.20E+2 | ±20% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
Calibrazione compatibile con la specifica GE Aviation P10TF3, accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0* | 2.81E-3 | ±0.45 decade or +182%, -65% |
| 1.0E-2 | 5.00E-3 | 1.50E-2 | ±5.00E-3 or ±50% |
| 5.0E-2 | 4.25E-2 | 5.75E-2 | ±15% |
| 1.0E-1 | 8.50E-2 | 1.15E-1 | ±15% |
| 5.0E-1 | 4.25E-1 | 5.75E-1 | ±15% |
| 1.0E+0 | 8.50E-1 | 1.15E+0 | ±15% |
| 1.0E+1 | 9.00E+0 | 1.10E+1 | ±10% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
ISO 17025 Accredited AMS 2769 Compatible Calibration
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-3 | 0.00E+0* | 5.00E-3 | ±4.00E-3 or ±400% |
| 1.0E-2 | 6.00E-3 | 1.40E-2 | ±40% |
| 5.0E-2 | 3.00E-2 | 7.00E-2 | ±40% |
| 1.0E-1 | 6.00E-2 | 1.40E-1 | ±40% |
| 5.0E-1 | 3.00E-1 | 7.00E-1 | ±40% |
| 1.0E+0 | 6.00E-1 | 1.40E+0 | ±40% |
| 1.0E+1 | 6.00E+0 | 1.40E+1 | ±40% |
| 1.0E+2 | 6.00E+1 | 1.40E+2 | ±40% |
| 7.6E+2 | 4.56E+2 | 1.00E+3* | ±40% |
Note: The calculated limit is negative, which isn't possible for absolute pressure, or is outside the gauge's measurement or display range, so the limit shown is the gauge's minimum or maximum reading.
7. 7B, 7E, 7M, 7F, 7FC, 7FCS Part 1
Applicable Controllers and Active Gauges
MX200, MM200, MC300, MP7ER, MP7FR
Calibrazione di fabbrica
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-5 | 6.31E-6 | 1.58E-5 | ±0.2 decade or -37%, +58% |
| 1.0E-4 | 6.31E-5 | 1.58E-4 | ±0.2 decade or -37%, +58% |
Calibrazione con tracciabilità NIST
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 6.31E-7* | 1.58E-6 | ±0.2 decade or -37%, +58% |
| 5.0E-6 | 3.16E-6 | 7.92E-6 | ±0.2 decade or -37%, +58% |
| 1.0E-5 | 6.31E-6 | 1.58E-5 | ±0.2 decade or -37%, +58% |
| 5.0E-5 | 3.16E-5 | 7.92E-5 | ±0.2 decade or -37%, +58% |
| 1.0E-4 | 6.31E-5 | 1.58E-4 | ±0.2 decade or -37%, +58% |
Calibrazione accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 6.31E-7* | 1.58E-6 | ±0.2 decade or -37%, +58% |
| 5.0E-6 | 3.16E-6 | 7.92E-6 | ±0.2 decade or -37%, +58% |
| 1.0E-5 | 6.31E-6 | 1.58E-5 | ±0.2 decade or -37%, +58% |
| 5.0E-5 | 3.16E-5 | 7.92E-5 | ±0.2 decade or -37%, +58% |
| 1.0E-4 | 6.31E-5 | 1.58E-4 | ±0.2 decade or -37%, +58% |
Calibrazione compatibile con la specifica GE Aviation P10TF3, accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 3.55E-7 | 2.81E-6 | ±0.45 decade or +182%, -65% |
| 5.0E-6 | 1.78E-6 | 1.40E-5 | ±0.45 decade or +182%, -65% |
| 1.0E-5 | 3.55E-6 | 2.81E-5 | ±0.45 decade or +182%, -65% |
| 5.0E-5 | 1.78E-5 | 1.40E-4 | ±0.45 decade or +182%, -65% |
| 1.0E-4 | 3.55E-5 | 2.81E-4 | ±0.45 decade or +182%, -65% |
ISO 17025 Accredited AMS 2769 Compatible Calibration
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 3.17E-7 | 3.16E-6 | ±0.5 decade or +216%, -68% |
| 5.0E-6 | 1.59E-6 | 1.58E-5 | ±0.5 decade or +216%, -68% |
| 1.0E-5 | 3.17E-6 | 3.16E-5 | ±0.5 decade or +216%, -68% |
| 5.0E-5 | 1.59E-5 | 1.58E-4 | ±0.5 decade or +216%, -68% |
| 1.0E-4 | 3.17E-5 | 3.16E-4 | ±0.5 decade or +216%, -68% |
Note: MM200 and MC300 tolerance at these points is 6.0E-7.
8. 7B, 7E, 7M, 7F, 7FC, 7FCS Part 2
Applicable Controllers and Active Gauges
Calibrazione di fabbrica
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-5 | 7.00E-6 | 1.30E-5 | ±30% |
| 1.0E-4 | 7.00E-5 | 1.30E-4 | ±30% |
Calibrazione con tracciabilità NIST
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 7.00E-7 | 1.30E-6 | ±30% |
| 5.0E-6 | 3.50E-6 | 6.50E-6 | ±30% |
| 1.0E-5 | 7.00E-6 | 1.30E-5 | ±30% |
| 5.0E-5 | 3.50E-5 | 6.50E-5 | ±30% |
| 1.0E-4 | 7.00E-5 | 1.30E-4 | ±30% |
Calibrazione accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 7.00E-7 | 1.30E-6 | ±30% |
| 5.0E-6 | 3.50E-6 | 6.50E-6 | ±30% |
| 1.0E-5 | 7.00E-6 | 1.30E-5 | ±30% |
| 5.0E-5 | 3.50E-5 | 6.50E-5 | ±30% |
| 1.0E-4 | 7.00E-5 | 1.30E-4 | ±30% |
Calibrazione compatibile con la specifica GE Aviation P10TF3, accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 3.55E-7 | 2.81E-6 | ±0.45 decade or +182%, -65% |
| 5.0E-6 | 1.78E-6 | 1.40E-5 | ±0.45 decade or +182%, -65% |
| 1.0E-5 | 3.55E-6 | 2.81E-5 | ±0.45 decade or +182%, -65% |
| 5.0E-5 | 1.78E-5 | 1.40E-4 | ±0.45 decade or +182%, -65% |
| 1.0E-4 | 3.55E-5 | 2.81E-4 | ±0.45 decade or +182%, -65% |
ISO 17025 Accredited AMS 2769 Compatible Calibration
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 3.17E-7 | 3.16E-6 | ±0.5 decade or +216%, -68% |
| 5.0E-6 | 1.59E-6 | 1.58E-5 | ±0.5 decade or +216%, -68% |
| 1.0E-5 | 3.17E-6 | 3.16E-5 | ±0.5 decade or +216%, -68% |
| 5.0E-5 | 1.59E-5 | 1.58E-4 | ±0.5 decade or +216%, -68% |
| 1.0E-4 | 3.17E-5 | 3.16E-4 | ±0.5 decade or +216%, -68% |
9. CC-10
Applicable Controllers and Active Gauges
Calibrazione di fabbrica
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-5 | 6.31E-6 | 1.58E-5 | ±0.2 decade or -37%, +58% |
| 1.0E+0 | 8.50E-1 | 1.15E+0 | ±15% |
| 7.6E+2 | 7.22E+2 | 7.98E+2 | ±5% |
Calibrazione con tracciabilità NIST
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 6.31E-7 | 1.58E-6 | ±0.2 decade or -37%, +58% |
| 5.0E-6 | 3.16E-6 | 7.92E-6 | ±0.2 decade or -37%, +58% |
| 1.0E-5 | 6.31E-6 | 1.58E-5 | ±0.2 decade or -37%, +58% |
| 5.0E-5 | 3.16E-5 | 7.92E-5 | ±0.2 decade or -37%, +58% |
| 1.0E-4 | 6.31E-5 | 1.58E-4 | ±0.2 decade or -37%, +58% |
| 1.0E-3 | 6.31E-4 | 1.58E-3 | ±0.2 decade or -37%, +58% |
| 1.0E-2 | 6.31E-3 | 1.58E-2 | ±0.2 decade or -37%, +58% |
| 1.0E-1 | 8.50E-2 | 1.15E-1 | ±15% |
| 1.0E+0 | 8.50E-1 | 1.15E+0 | ±15% |
| 1.0E+1 | 8.50E+0 | 1.15E+1 | ±15% |
| 1.0E+2 | 8.50E+1 | 1.15E+2 | ±15% |
| 7.6E+2 | 7.22E+2 | 7.98E+2 | ±5% |
Calibrazione accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 6.31E-7 | 1.58E-6 | ±0.2 decade or -37%, +58% |
| 5.0E-6 | 3.16E-6 | 7.92E-6 | ±0.2 decade or -37%, +58% |
| 1.0E-5 | 6.31E-6 | 1.58E-5 | ±0.2 decade or -37%, +58% |
| 5.0E-5 | 3.16E-5 | 7.92E-5 | ±0.2 decade or -37%, +58% |
| 1.0E-4 | 6.31E-5 | 1.58E-4 | ±0.2 decade or -37%, +58% |
| 1.0E-3 | 6.31E-4 | 1.58E-3 | ±0.2 decade or -37%, +58% |
| 1.0E-2 | 6.31E-3 | 1.58E-2 | ±0.2 decade or -37%, +58% |
| 1.0E-1 | 8.50E-2 | 1.15E-1 | ±15% |
| 1.0E+0 | 8.50E-1 | 1.15E+0 | ±15% |
| 1.0E+1 | 8.50E+0 | 1.15E+1 | ±15% |
| 1.0E+2 | 8.50E+1 | 1.15E+2 | ±15% |
| 7.6E+2 | 7.22E+2 | 7.98E+2 | ±5% |
Calibrazione compatibile con la specifica GE Aviation P10TF3, accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 3.55E-7 | 2.81E-6 | ±0.45 decade or +182%, -65% |
| 5.0E-6 | 1.78E-6 | 1.40E-5 | ±0.45 decade or +182%, -65% |
| 1.0E-5 | 3.55E-6 | 2.81E-5 | ±0.45 decade or +182%, -65% |
| 5.0E-5 | 1.78E-5 | 1.40E-4 | ±0.45 decade or +182%, -65% |
| 1.0E-4 | 3.55E-5 | 2.81E-4 | ±0.45 decade or +182%, -65% |
| 1.0E-3 | 3.55E-4 | 2.81E-3 | ±0.45 decade or +182%, -65% |
| 1.0E-2 | 5.00E-3 | 1.50E-2 | ±5.00E-3 or ±50% |
| 1.0E-1 | 8.50E-2 | 1.15E-1 | ±15% |
| 1.0E+0 | 8.50E-1 | 1.15E+0 | ±15% |
| 1.0E+1 | 9.00E+0 | 1.10E+1 | ±10% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
ISO 17025 Accredited AMS 2769 Compatible Calibration
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E-6 | 3.17E-7 | 3.16E-6 | ±0.5 decade or +216%, -68% |
| 5.0E-6 | 1.59E-6 | 1.58E-5 | ±0.5 decade or +216%, -68% |
| 1.0E-5 | 3.17E-6 | 3.16E-5 | ±0.5 decade or +216%, -68% |
| 5.0E-5 | 1.59E-5 | 1.58E-4 | ±0.5 decade or +216%, -68% |
| 1.0E-4 | 3.17E-5 | 3.16E-4 | ±0.5 decade or +216%, -68% |
| 1.0E-3 | 1.00E-9* | 5.00E-3 | ±4.00E-3 or ±400% |
| 1.0E-2 | 6.00E-3 | 1.40E-2 | ±40% |
| 1.0E-1 | 6.00E-2 | 1.40E-1 | ±40% |
| 1.0E+0 | 6.00E-1 | 1.40E+0 | ±40% |
| 1.0E+1 | 6.00E+0 | 1.40E+1 | ±40% |
| 1.0E+2 | 6.00E+1 | 1.40E+2 | ±40% |
| 7.6E+2 | 4.56E+2 | 1.00E+3* | ±40% |
Note: The calculated limit is negative, which isn't possible for absolute pressure, or is outside the gauge's measurement or display range, so the limit shown is the gauge's minimum or maximum reading.
10. 1E
Applicable Controllers and Active Gauges
Calibrazione di fabbrica
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 5.0E+0 | 4.00E+0 | 6.00E+0 | ±1.00E+0 or ±20% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 7.6E+2 | 7.22E+2 | 7.98E+2 | ±5% |
Calibrazione con tracciabilità NIST
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 5.0E+0 | 4.00E+0 | 6.00E+0 | ±1.00E+0 or ±20% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 4.0E+2 | 3.80E+2 | 4.20E+2 | ±5% |
| 7.6E+2 | 7.22E+2 | 7.98E+2 | ±5% |
Calibrazione accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E+0 | 0.00E+0 | 2.00E+0 | ±1.00E+0 or ±100% |
| 1.0E+1 | 9.00E+0 | 1.10E+1 | ±10% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 4.0E+2 | 3.80E+2 | 4.20E+2 | ±5% |
| 7.6E+2 | 7.22E+2 | 7.98E+2 | ±5% |
Calibrazione compatibile con la specifica GE Aviation P10TF3, accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E+0 | 8.50E-1 | 1.15E+0 | ±15% |
| 1.0E+1 | 9.00E+0 | 1.10E+1 | ±10% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 4.0E+2 | 3.60E+2 | 4.40E+2 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
ISO 17025 Accredited AMS 2769 Compatible Calibration
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E+0 | 6.00E-1 | 1.40E+0 | ±40% |
| 1.0E+1 | 6.00E+0 | 1.40E+1 | ±40% |
| 1.0E+2 | 6.00E+1 | 1.40E+2 | ±40% |
| 4.0E+2 | 2.40E+2 | 5.60E+2 | ±40% |
| 7.6E+2 | 4.56E+2 | 1.00E+3* | ±40% |
Note: The calculated limit is negative, which isn't possible for absolute pressure, or is outside the gauge's measurement or display range, so the limit shown is the gauge's minimum or maximum reading.
11. 1F
Applicable Controllers and Active Gauges
Calibrazione di fabbrica
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E+1 | 0.00E+0 | 2.00E+1 | ±1.00E+1 or ±100% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
Calibrazione con tracciabilità NIST
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E+1 | 0.00E+0 | 2.00E+1 | ±1.00E+1 or ±100% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 4.0E+2 | 3.60E+2 | 4.40E+2 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
Calibrazione accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E+1 | 0.00E+0 | 2.00E+1 | ±1.00E+1 or ±100% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 4.0E+2 | 3.60E+2 | 4.40E+2 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
Calibrazione compatibile con la specifica GE Aviation P10TF3, accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E+1 | 9.00E+0 | 1.10E+1 | ±10% |
| 1.0E+2 | 9.00E+1 | 1.10E+2 | ±10% |
| 4.0E+2 | 3.60E+2 | 4.40E+2 | ±10% |
| 7.6E+2 | 6.84E+2 | 8.36E+2 | ±10% |
ISO 17025 Accredited AMS 2769 Compatible Calibration
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 1.0E+1 | 6.00E+0 | 1.40E+1 | ±40% |
| 1.0E+2 | 6.00E+1 | 1.40E+2 | ±40% |
| 4.0E+2 | 2.40E+2 | 5.60E+2 | ±40% |
| 7.6E+2 | 4.56E+2 | 1.00E+3* | ±40% |
Note: The calculated limit is negative, which isn't possible for absolute pressure, or is outside the gauge's measurement or display range, so the limit shown is the gauge's minimum or maximum reading.
12. CDG (Capacitance Diaphragm Gauge, Capacitance Manometer)
Applicable Controllers and Active Gauges
Notes
FS indicates the full-scale range of the CDG (typically 0.1, 1, 10, 100, or 1000 Torr).
Calibrazione di fabbrica
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 0,1*FS | (0.1*FS) - 2% | (0,1*FS) + 2% | ±2% |
| 0,5*FS | (0.5*FS) - 2% | (0,5*FS) + 2% | ±2% |
| FS | FS - 2% | FS + 2% | ±2% |
Calibrazione con tracciabilità NIST
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 0,01*FS | (0.01*FS) - 10% | (0.01*FS) + 10% | ±10% |
| 0,1*FS | (0.1*FS) - 2% | (0,1*FS) + 2% | ±2% |
| 0,2*FS | (0.2*FS) - 2% | (0,2*FS) + 2% | ±2% |
| 0,3*FS | (0.3*FS) - 2% | (0,3*FS) + 2% | ±2% |
| 0,4*FS | (0.4*FS) - 2% | (0,4*FS) + 2% | ±2% |
| 0,5*FS | (0.5*FS) - 2% | (0,5*FS) + 2% | ±2% |
| 0,6*FS | (0.6*FS) - 2% | (0,6*FS) + 2% | ±2% |
| 0,7*FS | (0.7*FS) - 2% | (0,7*FS) + 2% | ±2% |
| 0,8*FS | (0.8*FS) - 2% | (0,8*FS) + 2% | ±2% |
| 0,9*FS | (0.9*FS) - 2% | (0,9*FS) + 2% | ±2% |
| FS | FS - 2% | FS + 2% | ±2% |
Calibrazione accreditata ISO 17025
| Punto di prova | Tolleranza inferiore | Tolleranza superiore | Tolleranza |
|---|---|---|---|
| 0,01*FS | (0.01*FS) - 10% | (0.01*FS) + 10% | ±10% |
| 0,1*FS | (0.1*FS) - 2% | (0,1*FS) + 2% | ±2% |
| 0,2*FS | (0.2*FS) - 2% | (0,2*FS) + 2% | ±2% |
| 0,3*FS | (0.3*FS) - 2% | (0,3*FS) + 2% | ±2% |
| 0,4*FS | (0.4*FS) - 2% | (0,4*FS) + 2% | ±2% |
| 0,5*FS | (0.5*FS) - 2% | (0,5*FS) + 2% | ±2% |
| 0,6*FS | (0.6*FS) - 2% | (0,6*FS) + 2% | ±2% |
| 0,7*FS | (0.7*FS) - 2% | (0,7*FS) + 2% | ±2% |
| 0,8*FS | (0.8*FS) - 2% | (0,8*FS) + 2% | ±2% |
| 0,9*FS | (0.9*FS) - 2% | (0,9*FS) + 2% | ±2% |
| FS | FS - 2% | FS + 2% | ±2% |
Calibrazione compatibile con la specifica GE Aviation P10TF3, accreditata ISO 17025
| Punto di prova | Tolleranza |
|---|---|
| 1.0E-6 ≤ Test Point < 1.0E-2 | ±0.45 decade or +182%, -65% |
| 1.0E-2 ≤ Test Point < 1.0E+1 | Greater of ±5.00E-3 or ±15% |
| 1.0E+1 ≤ Test Point ≤ 7.6E+2 | ±10% |
| 7.6E+2 < Test Point ≤ 1.0E+3 | ±5% |
ISO 17025 Accredited AMS 2769 Compatible Calibration
| Punto di prova | Tolleranza |
|---|---|
| 1.0E-6 ≤ Test Point < 1.0E-3 | ±0.5 decade or +216%, -68% |
| 1.0E-3 ≤ Test Point ≤ 1.0E+3 | Greater of ±4.00E-3 or ±40% |