Televac® AN 3011: MX200 and MM200 Analog Output Comparisons

1 Description

This application note shows the MX200 Vacuum Controller analog (0 to 10 V DC) output settings that are equivalent to an MM200 Vacuum Controller.

Back to Table of Contents

2 Comparison Chart

Green highlighted values are the default MX200 settings for that vacuum gauge type.

Red highlighted values are the default MM200 settings for that vacuum gauge type.

Vacuum Gauge Codice MM200 Impostazione MX200 Alto valore Basso valore
1E R1EN Lineare (LIN) 1,0E+03 Torr 1,0E+00 Torr
R1ELNon supportato--
1F R1EN Lineare (LIN) 1,0E+04 Torr 1,0E+01 Torr
R1ELNon supportato--
2A R2AN Lineare (LIN) 1,0E+00 Torr 1,0E-03 Torr
R2ALLogaritmico (LOG)1,0E+00 Torr1,0E-03 Torr
R2ATNon supportato--
R2ACNon supportato--
R2ARNon supportato--
4A R4AN Lineare per decennio (L-D) 1,0E+03 Torr 1,0E-03 Torr
R4AULineare (LIN)1,0E+00 Torr1,0E-03 Torr
R4ATNon supportato--
R4AL Logaritmico (LOG) 1,0E+03 Torr 1,0E-03 Torr
5A/5B/5C/5D/5E/5F R5AN Lineare (LIN) Fondo scala Fondo scala * 0,001
R5ALNon supportato--
7B R7BM Lineare per decennio (L-D) 1.0E-03 1.0E-07
R7BLLogaritmico (LOG)1.0E-031.0E-07
R7BENon supportato--
R7BN Lineare (LIN) 1.0E-03 1.0E-06
R7BANon supportato--
R7BTNon supportato--
7E R7EM Lineare per decennio 7E (L-D 7E) 1.0E-02 1.0E-08
R7EL Logaritmico (LOG) 1.0E-02 1.0E-11
R7ExLineare (LIN)1.0E-0x*1,0E-(x+3)*
7F/7FC/7FCS R7FM Lineare per decennio (L-D) 1.0E-02 1.0E-11
R7FL Logaritmico (LOG) 1.0E-02 1.0E-11
R7FxLineare (LIN)1.0E-0x*1,0E-(x+3)*

*Dependent on the value of x; 7E = 2≤x≤5, 7F/7FC/7FCS = 2≤x≤8.