1 Overview
The Televac® MX Series provides a modern replacement for legacy MP Series active vacuum gauges. This application note explains the differences in analog outputs, wiring, calibration, display, and set points that should be considered when replacing an MP2AR or MP4AR with an MX2A or MX4A.
Back to Table of Contents2 MP and MX Series Comparison
The table below summarizes the primary differences between the MX2A, MX4A, MP2AR, and MP4AR.
| Artículo de fondo | MX2A | MX4A | MP2AR | MP4AR |
|---|---|---|---|---|
| Rango | 1*10-4 a 1000 Torr | 1*10-4 a 1000 Torr | 1*10-3 a 20 Torr | 1*10-3 a 1000 Torr |
| Orientación |
Any Vertical |
Vertical | Cualquier | Vertical |
| Salidas analógicas | 7 seleccionables por el usuario | 7 seleccionables por el usuario | 2 de fábrica... | 3 de fábrica |
| Comunicaciones | RS-485 | RS-485 | Ninguno | Ninguno |
| Tiempo de respuesta | < 2 seconds | < 1 second | < 2 seconds | < 1 second |
3 Analog Output Compatibility
MP Series gauges provide fewer analog output options than MX Series gauges. The available output modes are summarized below.
| Series | Available Analog Outputs |
|---|---|
| MP Series | Log, Linear, Non-linear |
| MX Series | Log, Linear 1, Linear 2, Linear 3, Linear 4, Linear by decade, Non-linear |
MP2AR and MX2A Analog Output Comparison
| Salida del MP2AR | Salida equivalente a MX2A |
|---|---|
| Lineal | Lineal 4 |
| No lineal | No direct equivalent |
MP4AR and MX4A Analog Output Comparison
| Salida del MP4AR | Salida equivalente a MX4A |
|---|---|
| Bitácora | Bitácora |
| Lineal | Lineal 4 |
| No lineal | No lineal |
On MP Series gauges, analog output behavior is selected using jumpers and connector pin selection. On MX Series gauges, the analog output can be configured from the front panel or through RS-485 communications.
To configure an MX gauge to match the linear output of an MP gauge, select Linear 4 using the following steps:
- Navigate to the Measurement Screen.
- Press SEL four times. The screen will read Analog Out.
- Press ENT to unlock the screen.
- Press DOWN three times. The screen will read Linear 4.
- Press ENT to save the change.
4 Wiring and Pin Connections
The MP and MX Series use different connectors and pin assignments. Existing MP wiring must therefore be updated when installing an MX replacement.
Connector Comparison
| Conector MP | Conector MX |
|---|---|
| 9-pin D-sub | 15-pin D-sub |
Pin Mapping
| Conexión | MP Pin | MX Pin |
|---|---|---|
| Potencia (+) | Pin 3 | Pin 8 |
| Poder (-) | Pin 4 | Pin 15 |
| Registro/Salida analógica lineal (+) | Pin 5 | Pin 9 |
| Salida analógica no lineal (+) | Pin 7 | Pin 9 |
| Salida analógica (-) | Pin 8 | Pin 1 |
| Relevo de punto de ajuste NO | Pin 1 | Pin 4 |
| Relevo de punto de ajuste NC | Pin 2 | Pin 3 |
| Punto de ajuste Relé C | Pin 6 | Pin 11 |
Overall Pin Diagram Comparison
5 Calibration
Calibration points on MP Series gauges are adjusted using potentiometers (pots). MX Series calibration is performed through the front panel or RS-485 communications.
Puntos de calibración de MP2AR y MX2A
| Punto de calibración | MP2AR | MX2A |
|---|---|---|
| Cero | Sí | Sí |
| 10 Torr | No | Sí |
| 760 Torr (Atmósfera) | No | Sí |
Puntos de calibración de MP4AR y MX4A
| Punto de calibración | MP4AR | MX4A |
|---|---|---|
| Cero | Sí | Sí |
| 1 Torr | No | Sí |
| 760 Torr (Atmósfera) | Sí | Sí |
The additional calibration points available on MX Series gauges allow calibration to be performed at more points across the measurement range.
Back to Table of Contents6 Display
MX Series gauges use a higher-resolution front-facing OLED display, while MP Series gauges use a top-mounted 3-character LED display.
| Descripción | Pantalla MP | Pantalla MX |
|---|---|---|
| Escriba | 3-character 7-segment LED | 256 × 64 color OLED |
| Ubicación | Top | Frente |
| Distancia de lectura | 3 m (10 pies) | 3 m (10 pies) |
The MX display also provides access to configuration settings directly from the front panel.
Back to Table of Contents7 Set Points
MX Series gauges provide two set points and support higher switching loads than MP Series gauges.
| Descripción | MP | MX |
|---|---|---|
| Puntos de ajuste | 1 | 2 |
| Escriba | Relevo | Relevo, Colector Abierto |
| Clasificación de los contactos de los relés | 0.5 A @ 24 V DC | 2 A @ 220 V DC (250 V AC) |
| Clasificación de coleccionista abierto | Ninguno | 2 A @ 60 V DC |
8 RS-485 Communications
MX Series gauges include RS-485 communications, while MP Series gauges do not. RS-485 allows multiple addressable gauges to share the same communications bus and enables remote configuration without accessing the gauge's front panel.