1 Description
This application note shows the MX200 Vacuum Controller analog (0 to 10 V DC) output settings that are equivalent to an MM200 Vacuum Controller.
Back to Table of Contents2 Comparison Chart
Green highlighted values are the default MX200 settings for that vacuum gauge type.
Red highlighted values are the default MM200 settings for that vacuum gauge type.
| Vacuum Gauge | Código MM200 | Configuración MX200 | Alto valor | Bajo valor... |
|---|---|---|---|---|
| 1E | R1EN | Lineal (LIN) | 1,0E+03 Torr | 1,0E+00 Torr |
| R1EL | No se admite | - | - | |
| 1F | R1EN | Lineal (LIN) | 1,0E+04 Torr | 1,0E+01 Torr |
| R1EL | No se admite | - | - | |
| 2A | R2AN | Lineal (LIN) | 1,0E+00 Torr | 1,0E-03 Torr |
| R2AL | Logarítmico (LOG) | 1,0E+00 Torr | 1,0E-03 Torr | |
| R2AT | No se admite | - | - | |
| R2AC | No se admite | - | - | |
| R2AR | No se admite | - | - | |
| 4A | R4AN | Lineal por décadas (L-D) | 1,0E+03 Torr | 1,0E-03 Torr |
| R4AU | Lineal (LIN) | 1,0E+00 Torr | 1,0E-03 Torr | |
| R4AT | No se admite | - | - | |
| R4AL | Logarítmico (LOG) | 1,0E+03 Torr | 1,0E-03 Torr | |
| 5A/5B/5C/5D/5E/5F | R5AN | Lineal (LIN) | Escala completa | Escala completa * 0,001 |
| R5AL | No se admite | - | - | |
| 7B | R7BM | Lineal por décadas (L-D) | 1.0E-03 | 1.0E-07 |
| R7BL | Logarítmico (LOG) | 1.0E-03 | 1.0E-07 | |
| R7BE | No se admite | - | - | |
| R7BN | Lineal (LIN) | 1.0E-03 | 1.0E-06 | |
| R7BA | No se admite | - | - | |
| R7BT | No se admite | - | - | |
| 7E | R7EM | Lineal por décadas 7E (L-D 7E) | 1.0E-02 | 1.0E-08 |
| R7EL | Logarítmico (LOG) | 1.0E-02 | 1.0E-11 | |
| R7Ex | Lineal (LIN) | 1.0E-0x* | 1.0E-(x+3)* | |
| 7F/7FC/7FCS | R7FM | Lineal por décadas (L-D) | 1.0E-02 | 1.0E-11 |
| R7FL | Logarítmico (LOG) | 1.0E-02 | 1.0E-11 | |
| R7Fx | Lineal (LIN) | 1.0E-0x* | 1.0E-(x+3)* |
*Dependent on the value of x; 7E = 2≤x≤5, 7F/7FC/7FCS = 2≤x≤8.
Back to Table of Contents